Lifeline & Reviver View AED (2020-04-20)
Press Release
By:
Ref: Health Canada
Published: 04/29/2020
Lifeline & Reviver View AED (2020-04-20) - Recalls and safety alerts
Starting date:
April 20, 2020
Posting date:
April 29, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72893
Last updated:
2020-04-29
A. Lifeline View AED
B. Reviver View AED
As part of the manufacturing process, 4 DDU-2000 units AEDs failed quality tests. Investigation demonstrated that a commercial off-the-shelf (COTS) component with thermal sensitivity within the AED circuitry created a reboot behavior causing error codes to occur during test. This behavior may result in failure to deliver shock and/or failure to deliver shock in range.
- Serial number 400131246
- Serial number 400131264
- DDU-E2300
Manufacturer
Defibtech, LLC 741 Boston Post, Suite 201 Guilford 6437 Connecticut UNITED STATES
- Serial number 400128145
- DDU-C2300
Manufacturer
Defibtech, LLC 741 Boston Post, Suite 201 Guilford 6437 Connecticut UNITED STATES